Section 1
1-1
General
Service Manual Purpose
The purpose of this manual is to supply information to service personnel so they can maintain the ELI 150
Interpretive and Non-Interpretive Electrocardiograph at the assembly and subassembly level. Although the manual
includes parts lists, mechanical assembly parts, and printed circuit board information, it is intended to function
primarily as a guide to preventative and corrective maintenance and electrical repairs considered field repairable.
User Safety Information
Periodic Safety Inspections
Follow the recommended maintenance schedule. Inspect the power cord and transmission cables
periodically for fraying or other damage and replace as needed. Broken or frayed wires may cause
interference or loss of signal. Pay particular attention to points where wires enter connectors.
Proper Power Cord
Use only the power cord specified for the equipment. This product requires a three-wire, (18 gauge, SJT-
grade) power cord, which is supplied with a three-terminal, polarized plug (hospital grade) for connection
to the power source and protective ground. Use only a power outlet with a protective ground outlet. An
interruption of the grounding connection could cause an electrical shock hazard.
Proper Fuse
Use only the fuse specified for the equipment (identical in type, voltage and current rating). Substituting a
different fuse type could cause a fire hazard. Always make sure fuses have been installed before operating
the unit.
Do Not Operate in Explosive Atmospheres
Do not operate the ELI 150 in the presence of flammable gasses or anesthetics; this environment could
cause an explosion. Refer to Operator's Manual Safety Information: Warning(s) and Caution(s).
Use Only Safe Methods of Interconnection
To prevent electrical shock from the product when it is connected to other electrical equipment, proper
grounding is essential. Refer to Operator's Manual Safety Information: Warning(s) Peripheral equipment.
Do Not Mount Product above Patient
Do not mount or place the product where it could fall on a patient or where it could be accidentally knocked
off a shelf or other mounting arrangement.
Recommended Accessories
For the patient's safety and optimum equipment performance, use only the accessories specified by Mortara
Instrument, Inc.
1
Содержание ELI 150
Страница 3: ......
Страница 10: ...viii...
Страница 14: ...xii...
Страница 36: ......
Страница 41: ...Section 4 4 5 Open the writer door by lifting up on the writer door latch and sliding the door to the left...
Страница 44: ...ELI 150 4 8 Re assemble in reverse order...
Страница 54: ...ELI 150 4 18 Re assemble in reverse order...
Страница 72: ......
Страница 74: ...ELI 150 5 2 ELI 150 Block Diagram...
Страница 77: ...Section 5 5 5 Specifications...
Страница 80: ......
Страница 100: ......