Directions for Use
Chapter 3 Calibrating the Spirometer
25
Figure 10. Calibration Code on Flow Transducer Package Label
3.
Press the desired softkey, as listed here.
•
Verify Calibration
To verify the accuracy of the system (without recalibrating). None of the
calibration factors will be saved. The date that prints on reports will be the last
calibration date, not the verification date.
•
Calibrate 1 Flow
To calibrate the system using one inhale/exhale cycle.
•
Calibrate 3 Flows
To calibrate the system using three inhale/exhale cycles at three different rates.
The “attach flow transducer” prompt appears, as shown in
.
Figure 11. “Attach Flow Transducer” Prompt
D
r
awi
n
g No.
3
0015257 VER. F
8
LOT
2
QTY 100
Disposable Flow T
r
a
n
sd
u
ce
r
s
CALIBRATION CODE WKKVDXPB7
0050
200x-xx
4
3
41 State St
r
eet Road
Ska
n
eateles Falls, NY 1
3
15
3
USA
www. welchally
n
.com
REF 70
3
419
Transducer
“cal” code
Lot code
Spirometer Calibration
9:17AM Oct 16 08
Attach the flow transducer
to the syringe,
Pull the plunger out,
Then select continue
Back
Continue
At any time, you can press
Back
to return to
the initial calibration screen, as shown in
Содержание CP 200
Страница 1: ...CP 200 Spirometry Option Directions for Use ...
Страница 34: ...30 Chapter 3 Calibrating the Spirometer Welch Allyn CP 200 Spirometry Option ...
Страница 52: ...48 Chapter 4 Performing Spirometry Tests Welch Allyn CP 200 Spirometry Option ...
Страница 53: ...49 5 Troubleshooting Problem Solving Suggestions 50 Limited Warranty 51 Service Policy 51 ...
Страница 56: ...52 Chapter 5 Troubleshooting Welch Allyn CP 200 Spirometry Option ...
Страница 80: ...76 Appendix D Predictive Norms etc Welch Allyn CP 200 Spirometry Option ...
Страница 91: ......