connection to the host system in Advanced settings. Furthermore, you can add patients
to a patient list on the monitor and then send them to the host system, and you can also
retrieve patient lists from a host system. In the Intervals Monitoring and Spot Check
profiles, retrieving the patient list requires user action. In the Continuous Monitoring
profile, the patient list on the monitor updates automatically when the monitor is
connected to the host system.
Once you assign a patient to a monitor and send this information to the host system, you
can edit patient information on the monitor as noted in the following table:
Editable patient information based on profile
and host system communication
Patient ID
Patient name
Patient
location
Patient type
Spot Check and Intervals profiles
X
X
X
X
Continuous Monitoring profile, communicating with
the host system
X
X
Continuous Monitoring profile, not communicating
with the host system
X
X
X
Note
In many cases, changing patient information opens an onscreen dialog
asking you to confirm the action. These dialogs appear whenever the action
results in clearing patient measurement data or patient context.
If a monitor loses its connection to the host system, the device continues to monitor the
patient, display patient data, and generate alarms and messages. When connectivity is
restored, the monitor resumes sending vital-signs trends and episodic data.
Monitor tab
The Monitor tab provides the following controls related to continuous patient monitoring
at the device and the streaming of patient vital-signs data to a host system:
•
End Monitoring – Stop continuous monitoring on the current patient and clear patient
data on the device.
•
Pause – Pause continuous monitoring as well as alarms for a specified period and
retain patient data on the device.
•
Change pause interval – Increase or decrease the pause interval in 15 minute
increments up to 2 hours.
•
Connect to host system – Manually connect to a host system and start or resume
the streaming of patient vital-signs data when the device does not connect
automatically.
•
Disconnect from host system – Manually disconnect the device from a host system
yet continue monitoring the patient and retain patient vital-signs data on the device.
This control is only available after connecting to a host system.
The sections which follow provide instructions for using these controls.
Note
This tab is visible only when the Continuous Monitoring profile is active. It
is not available in the Office, Spot Check, or Intervals Monitoring profiles.
Note
The Continuous Monitoring profile is available only on devices with a
Continuous Monitoring profile license.
74 Host system
Welch Allyn Connex
®
Devices
Содержание Connex IWS
Страница 1: ...Welch Allyn Connex Devices Directions for use Software version 2 3X...
Страница 10: ...4 Introduction Welch Allyn Connex Devices...
Страница 14: ...8 Symbols Welch Allyn Connex Devices...
Страница 20: ...14 Screen elements Welch Allyn Connex Devices...
Страница 28: ...22 About warnings and cautions Welch Allyn Connex Devices...
Страница 72: ...66 Navigation Welch Allyn Connex Devices...
Страница 96: ...90 Profiles Welch Allyn Connex Devices...
Страница 106: ...100 Patient data management Welch Allyn Connex Devices...
Страница 120: ...114 Alarms Welch Allyn Connex Devices...
Страница 184: ...178 Patient monitoring Welch Allyn Connex Devices...
Страница 212: ...206 Advanced settings Welch Allyn Connex Devices...
Страница 244: ...238 Specifications Welch Allyn Connex Devices...
Страница 250: ...244 Standards and compliance Welch Allyn Connex Devices...
Страница 252: ...246 Guidance and manufacturer s declaration Welch Allyn Connex Devices...
Страница 266: ......