Welch Allyn

®

AED 10

TM

 Trainer

Operating Instructions

Software version 01.04.XX

Содержание AED 10

Страница 1: ...Welch Allyn AED 10 TM Trainer Operating Instructions Software version 01 04 XX...

Страница 2: ...re may not be copied decompiled reverse engineered disassembled or otherwise reduced to human perceivable form This is not a sale of the software or any copy of the software all right title and owners...

Страница 3: ...l accurately mimic the operation of the defibrillator and provide a realistic demonstration Twelve different scenarios are available Multiple language selections are available Hand held IR remote cont...

Страница 4: ...urrent battery level The Trainer displays text and voice prompts when the battery needs to be changed Note The Battery Level Indicator is only available on Welch Allyn AED 10 defibrillators using soft...

Страница 5: ...ng a scenario and continues after the scenario is resumed Battery Level While in Trainer mode the Trainer graphically displays the battery level in the upper middle portion of the screen The battery l...

Страница 6: ...ior to entering the Volume menu and returns to the Main Menu Contrast Pressing the Enter button while the Contrast menu item is highlighted displays the Contrast menu allowing the user to select a con...

Страница 7: ...splays the CPR menu with CPR Timer highlighted Pressing the Next button highlights CPR Tempo and pressing Next again highlights Back With CPR Timer highlighted pressing the Enter button displays the C...

Страница 8: ...f CPR chest compressions per minute options are 90 95 100 105 110 115 120 or Disable The Save menu item saves the currently selected value and returns to the CPR Menu The Cancel menu item restores CPR...

Страница 9: ...Defaults Pressing the Enter button while the Restore Default menu item is highlighted displays the restore defaults menu Pressing the Next button scrolls down highlighting the next menu item If the E...

Страница 10: ...he Enter button while the Maintenance menu item is highlighted displays the Maintenance Menu This menu is for factory use only Maintenance Service System Test Enter Enter More Menu More Menu Maintenan...

Страница 11: ...hm CPR 3 Ventricular fibrillation with two shock conversion Shockable rhythm 2 Shocks Non shockable rhythm CPR Non shockable rhythm CPR 4 Non shockable rhythm Non shockable rhythm CPR Non shockable rh...

Страница 12: ...it will flash to indicate that a command has been received The status indicator LED will flash continuously when a scenario has been paused Refer to the diagram and descriptions listed below Table 3...

Страница 13: ...the Instructor to select Ventricular Fibrillation with one shock conversion followed by refibrillation The Trainer will automatically turn off after the selection has been made Note Scenarios 11 and 1...

Страница 14: ...trol Approximately two years Number of training scenarios 12 Storage temperature without batteries 22 F to 158 F 30 C to 70 C Operating temperature 32 F to 122 F 0 C to 50 C Weight without batteries 1...

Страница 15: ...idity should be at least 30 Electrical fast transient burst IEC 61000 4 4 2 kV for power supply lines 1 kV for input output lines Not applicable Not applicable Surge IEC 61000 4 5 1 kV differential mo...

Страница 16: ...the frequency range 80 MHz to 2 5 GHz are intended to decrease the likelihood that mobile portable communications equipment could cause interference if it is inadvertently brought into patient areas F...

Страница 17: ...79 3 79 7 27 100 11 70 12 00 12 00 23 00 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be determined using the equation applic...

Страница 18: ...16 Welch Allyn AED 10 Trainer...