Emissions and immunity information
Electromagnetic emissions
The Welch Allyn Welch Allyn Home Blood Pressure Monitor RPM-BP100SBP (Welch Allyn Home Blood Pressure Monitor)
is intended for use in the electromagnetic environment specified below. The customer or user of the Welch Allyn Home
Blood Pressure Monitor should assure that it is used in such an environment.
This equipment needs to be installed and put into service in accordance with the information provided in the user
manual.
CAUTION
Wireless communications equipment, such as wireless home network devices, mobile
phones,cordless telephones and their base stations, walkie-talkies can affect this equipment and
should be kept at least a distance d=3,3m away from the equipment.
Note
As indicated in Table 6 of IEC 60601-1-2:2007 for ME Equipment, a typical cell phone with a maximum
output power of 2W yields d=3,3m at an IMMUNITY LEVEL of 3V/m.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The Welch Allyn Home Blood Pressure Monitor uses RF energy only for
its internal function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The Welch Allyn Home Blood Pressure Monitor is suitable for use in all
establishments other than domestic establishments and those directly
connected to the public low-voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
Electromagnetic immunity
The Welch Allyn Home Blood Pressure Monitor is intended for use in the electromagnetic environment specified below.
The customer or the user of the Welch Allyn Home Blood Pressure Monitor should assure that it is used in such an
environment.
Immunity test
IEC 60601 test level Compliance level
Electromagnetic environment - guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
±8kV contact
±15kV air
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power supply
lines
±1 kV for input/output
lines
±2 kV for power supply
lines
Mains power quality should be that of a typical
commercial or hospital environment.
Surge
IEC 61000-4-5
±1 kV line(s) to line(s)
±2 kV line(s) to earth
±1 kV differential mode Mains power quality should be that of a typical
commercial or hospital environment.
Directions for use
Appendix 25