3
19
Internal Configuration
You can change several device operating parameters in the Internal Configuration Mode.
When changed, these settings become the default power-up settings. You will also see
non-changeable device configurations for technical service purposes.
To enter the Internal Configuration Mode:
1.
Turn the Spot Vital Signs off.
2.
Press and hold the
Power
and
Blood Pressure Start/Stop
buttons. The device
enters the Internal Configuration Mode and displays the software version.
3.
Press the
Mode
button to cycle through the Internal Configuration menu until you see
the menu option displayed on the screen.
4.
Use the
Next Patient/Clear/Cancel
or
Blood Pressure Start/Stop
buttons to
change the default setting.
5.
Press the
Mode
button once to save the change and press the
Power
button to exit
the Internal Configuration Mode.
Table 2. Configuration Menu Options
Setting
Description
Blood Pressure Calibration
Displays “Cal”
Prepares the Spot Vital Signs for calibration. Only qualified personnel should verify the
Spot Vital Signs blood pressure calibration. For more details, see “Checking the Blood
Pressure Calibration” on page 26.
Inflation Pressure Preset Level
Displays “PrP”
120, 140, 160, 180, 200, 240, 280 mmHg. Factory default is 160 mmHg.
Pressure Preset Level
Displays “PrP“
On or off. Disables or enables the front panel Pressure Preset button.
Backlight
Displays “BLT”
On or off.
Mean Arterial Pressure
Displays “MAP”
On or off.
Date/Time
Changes or updates the actual date and time.
Temperature Scale
Displays “TMP MOD”
Fahrenheit (°F) or Celsius (°C) Normal Mode / Fahrenheit (°F) or Celsius (°C) Monitor
Mode
Blood Pressure Units
Displays “BP”
mmHg or kPa.
Battery Readings
Displays “BAT”
Displays the total battery voltage.
Battery Life
Displays “LFE”
Total number unit measurements. Displayed information only; operator cannot change.
Содержание 4200B
Страница 3: ...Welch Allyn Spot Vital Signs Directions for Use...
Страница 6: ...iv Welch Allyn Spot Vital Signs...
Страница 22: ...14 Welch Allyn Spot Vital Signs...
Страница 28: ...20 Internal Configuration Welch Allyn Spot Vital Signs...
Страница 36: ...28 Welch Allyn Spot Vital Signs...
Страница 42: ...34 Welch Allyn Spot Vital Signs...
Страница 44: ...36 Welch Allyn Spot Vital Signs...
Страница 50: ...42 Welch Allyn Spot Vital Signs...
Страница 60: ...52 Welch Allyn Spot Vital Signs...
Страница 70: ...62 Welch Allyn Spot Vital Signs...
Страница 72: ...64 Warranty Welch Allyn Spot Vital Signs...
Страница 73: ......
Страница 74: ...Reorder No 4200 87E Material No 706272 Ver C...