70
EN
Fine dust filter
1.
Remove the coarse dust filter as
described above.
2.
Remove the fine filter.
3.
Check the soiling level of the filter.
The fine filter must be changed when it has
turned dark grey in colour, however not lat-
er than 250 operating hours. In this case the
filter change indicator
will appear.
4.
Insert the filters and filter compartment
cover in reverse order.
5.
To cancel the filter change indicator
hold down the on / off switch
switching on the unit until the indicator
is extinguished.
If the fine filter is changed before the 250
operating hours have elapsed due to soiling,
the hours meter has to be set to zero:
6.
To do so hold down the on / off
switch
when switching on the unit.
The filter change indicator lights up af-
ter 3 approx. seconds and is extin-
guished after another 3 seconds. Only
then release the switch.
Bacteria filter
When changing the bacteria filter, please
observe the relevant operating instructions.
9.3
Change pressure
measurement tube
1.
Detach the sleeve on the creased hose
from the adapter
2.
Remove the pressure measurement
tube
3.
Remove the pressure measurement
tube from the adapter