5. Operation
5.1
Automatic on / off
switching system
Activate automatic system
A0
= Automatic system not activated
Automatiken ej i funktion
Automatik ikke aktiveret
A1
= Automatic system activated
Automatiken i funktion
Automatik aktiveret
1.
With the SOMNO
vent
S switched off
press the softstart button
until the cur-
rent setting (
A0
) appears in the display.
2.
the setting to
A1
.
If a button is not operated for 5 seconds, the
display will disappear and "
0
" will reappear.
When the automatic system is activated,
you can switch SOMNO
vent
S on and off
using the on / off switch
or
using the
pressure available in the mask.
•
SOMNO
vent
S is switched on automat-
ically as soon as the patient breathes
into the mask (pressure > 0.5 mbar).
•
If the mask is removed, the unit will
switch off after 5 seconds.
Switch off automatic system
A0
= Automatic system not activated
Automatiken ej i funktion
Automatik ikke aktiveret
A1
= Automatic system activated
Automatiken i funktion
Automatik aktiveret
1.
With the SOMNO
vent
S switched off
press the softstart button
until the
current setting (
A1
) appears in the dis-
play.
2.
the setting to
A0
.
If a button is not operated for 5 seconds, the
display will disappear and "
0
" will reappear.
When the automatic system is switched off,
you can only switch SOMNO
vent
S on and
off using the on / off switch
If the unit is not used for 15 minutes, it
switches off automatically.