14 Appendix
14.1.10 Factory settings for emergency modes and
ventilation modes
Ventilation
parameters
Adult
Child
Infant
Pressure ramp
Averaging
Averaging
Averaging
ExTr
35 % Flow
Level 2
35 % Flow
Level 2
35 % Flow
Level 2
Freq.
10/min
20/min
30/min
I:E
1:1.7
1:1.7
1:2.8
InTr
5 l/min
Level 2
5 l/min
Level 2
5 l/min
Level 2
PEEP
0 mbar
0 mbar
0 mbar
PEEP CCSV
3 mbar
3 mbar
–
pInsp
20 mbar
15 mbar
15 mbar
pInsp CCSV
60 mbar
40 mbar
–
pMax
30 mbar
25 mbar
20 mbar
pMax CPR Contin.
30 mbar
25 mbar
20 mbar
pMax CPR Man.
20 mbar
20 mbar
20 mbar
pMax RSI
30 mbar
25 mbar
20 mbar
Trigger window
20 % Te
20 % Te
20 % Te
Vt
500 ml
200 ml
60 ml
Δ
pASB
0 mbar
0 mbar
0 mbar
Alarm limit
Adult
Child
Infant
MVe
6.5 l/min
5.2 l/min
2.3 l/min
MVe
3.5 l/min
2.8 l/min
1.3 l/min
f
13/min
26/min
39/min
Apnea
30 s
30 s
30 s
etCO
2
49 mmHg/6 vol%
49 mmHg/6 vol%
49 mmHg/6 vol%
etCO
2
27 mmHg/4 vol%
27 mmHg/4 vol%
27 mmHg/4 vol%