WM
6807
1
0
5
/2
021
MEDUMAT Standard
2
EN
181
8 Hygienic reprocessing
8.5 Cleaning parts manually
The hygienic reprocessing is prepared
.
1. The parts approved for manual cleaning can be found in the
cleaning and disinfection plan
.
2. The cleaning agents, doses and exposure time for the individual
parts can be found in the cleaning and disinfection plan
(see "8.11 Cleaning and disinfection plan", page 196)
.
3. Prepare the cleaning solution in accordance with the
specifications of the cleaning agent manufacturer.
4. To remove all visible dirt: Thoroughly brush down parts on the
inside and outside using a standard soft brush that is suitable
for use on plastics and that has been wetted with the cleaning
agent.
When doing so, note:
•
Keep uneven surfaces and grooves (e.g., top and bottom of
MEDUtrigger, adjusting knob, connection for ventilation
hose) moist for the duration of the exposure time and
brush them off particularly thoroughly (e.g., with the brush
with article number 09084 from Interlock Medizintechnik
GmbH).
•
Brush off the inner lumen of the hoses/tubes with a special
lumen brush.
Содержание MEDUMAT Standard2
Страница 1: ...MEDUMAT Standard2 Ventilator Instructions for Use for Devices from Software Version 5 1...
Страница 110: ...110 EN MEDUMAT Standard2 WM 68071 05 2021 5 User menu 5 3 Settings in the user menu 5 3 1 Mode 5 2 Submenu Mode...
Страница 269: ......
Страница 270: ......
Страница 271: ......