Changing the

 

ventilation mode

Requirements:
•  The device is switched on.
•  A ventilation mode is set.
1. Briefly press the menu button 

.

2. Select the 

Mode

 field. 

3. Select a ventilation mode.
4. Adjust the ventilation parameters and press 

start

.

Ventilating the patient

Selecting emergency mode

1. Briefly press the On/Off button 

.

2. Select emergency mode:
•  Emergency Pediatric
•  Emergency Child
•  Emergency Adult

Selecting previous patient

1. Briefly press the On/Off button 

.

2. Select 

Previous patient

 field.

RSI mode

Performing RSI ventilation

Requirements:
•  The device is switched on.
•  The MEDUtrigger is connected to the device.
• 

RSI

 was set as the ventilation mode for a 

  new patient. The Demand function is 
  automatically activated.

1. For the function 

Manual

, navigate to the 

  field 

Demand

.

2. Select the 

Manual

 field.

3. To perform continuous ventilation 
  following successful airway management, 
  select the 

Contin.

 field. The device 

  switches to IPPV mode.

Selecting a new patient

1. Briefly press the On/Off button 

.

2. Select 

New patient

 field.

3. Select height and gender 

or

 navigate to the  

 

Height

 field and turn the navigation knob further  

  to select the desired patient group.
4. Press the navigation knob 

next

.

5. Select a ventilation mode.
6. If necessary: Adjust the ventilation parameters and  
  press 

start

.

Pausing the ventilation

Requirements:
•  Cardiac rhythm analysis or defibrillation  

 

  is being performed. 
•  The device is switched on.
•  The CPR mode is set.
•  Continuous ventilation
 

 is activated.

1. Select the Pause 

 field.

  The ventilation is paused for 50 seconds.

CPR mode

Performing CPR ventilation

Requirements:
•  The device is switched on.
•  The MEDUtrigger is connected to the device.
•  A ventilation mode for a new patient is set.

1. Briefly press the CPR button 

.

  The device switches to the mode CPR 30:2/CPR 15:2/CPR IPPV (depending on    
  the preset). The green LEDs on the MEDUtrigger light up.

2. If necessary:
  Change algorithm:
  • 30:2 
  • 15:2
  • 

 Continuous

3. Press and hold the MEDUtrigger button 
  during the ventilation phase until two
  mechanical breaths are performed.

4. To exit the CPR mode,
  press the CPR button 

 again.

To enable the selection of the function 

Manual

, a MEDUtrigger must 

be connected. Otherwise, this function will not be displayed.

If you do not select a menu within 20 seconds, the device automati-

cally switches to the 

Previous patient

 menu and immediately begins 

ventilation of the patient. In this case, the device uses the ventilation mode 

and ventilation parameters set for the last patient to be ventilated. To switch 

off  the  countdown  before  the  20  seconds  have  elapsed,  move  one  of  the 

navigation knobs.