Final check
15
3.6 Function check of controls (key test)
1. Provide a pressurized gas supply to MEDUMAT Easy CPR and to the flow meter.
2. Open oxygen cylinder.
3. Switch on device.
4. Switch from mask ventilation to tube ventilation.
5. Press the alarm mute button
6. Press the CPR key
.
Requirement:
All the controls must work perfectly.
3.7 Function check and alarms
1. Provide a pressurized gas supply to MEDUMAT Easy CPR and to the flow meter.
2. Open oxygen cylinder.
3. Switch on device.
13
14