Technical data
65
EN
12.1 Pneumatic / electronic systems
V6
V5
V4
V2
V1
V3
Relief valve
3/2-way valve
Function
Pneumatic system
Pressure regulator
Booster valve with
Flow
Coupling element
Respiratory frequency
Position
Demandflow
Battery
Sound
source
M/T
O/I
LED bar graph
4bar
Electronic
system
2,7-6bar
2,5 bar
RF/MV
100mbar
Patient valve
Exp.
Insp.
metering facility
100mbar10mbar
Pp
E
Pp
E
Pp
E
MPU speech module
Stenosis
Disconnection
< 2,7
Batterie
P
Ventilation
hose
Pressure
gauge tube
The input pressure at p is max. 6 bar. This is reduced
by V1 to 2.5 bar dyn. This is the input pressure at V2,
V3 and V4.
Inspiration
An electrical impulse to V2 opens V3 and closes V4.
Oxygen flows through the ventilation hose to the pa-
tient valve. If the ventilation pressure in the patient
valve reaches >100 mbar, the relief valve V5 will
open.
Expiration
A fresh electrical impulse closes V2. The relief valve
V4 opens and vents the ventilation hose. The patient
breathes out through the patient valve.