Technical data
EN
11.2 Pneumatic diagram
11.3 Safety distances
Recommended safety distances between portable and mobile
HF telecommunications equipment (e.g. mobile phones) and
CPAP 20e or CLICK 2
Nominal output
of HF equipment
in W
Safety distance dependent on transmission frequency
in m
150 kHz - 80 MHz
80 MHz - 800 MHz
800 MHz – 2.5 GHz
0.01
0.12
0.12
0.24
0.1
0.37
0.37
0.74
1
1.17
1.17
2.34
10
3.69
3.69
7.38
100
11.67
11.67
23.33
O
2
pressure source
(max. 1.5 bar)
Filter
Optional
bacteria filter
Breathing tube
1.8 m
Exhalation
system
Pressure sensor for
patient pressure
Patient mask
(nasal or
full-face)
Ambient
air inlet
Fan
Optional
humidifier
Flow regulator
to max. 4 l/min
Optional
O
2
valve