CPR Touch XL, Control Unit
Index: 03 Date modified: 22/06/2022
OI No.: BA MR 001-03 Steuereinheit CPR Touch XL EN.docx
Page 57 of 70
9.2.1
Wiring diagram power pack for device
2S
DANGER!
Risk of death due to high voltage. All electrical work on the device
must only be carried out by trained specialists in accordance with
the applicable safety regulations! Internal fuses may only be
renewed once the voltage has been disconnected and secured
against being reactivated!
Fuse in the control housing (accessible from the outside)
Fuses
Current
Fuse type
Function
6,3 A
5 x 20 mm
Primary fuse
mains lead
Fuses on the power pack
Fuses
Current
Fuse-type
Function
F1
315 mA
Micro-fuse
24 VDC sensors
F2
2 A
Micro-fuse
Primary fuse
power pack
F3
1 A
Micro-fuse
with the CPR Touch XL 2S without function
F4
4 A
Micro-fuse
230 VDC output relays K2 and K3
F5
1 A
Micro-fuse
with the CPR Touch XL 2S without function