![Waygate Technologies Krautkramer USM Go Скачать руководство пользователя страница 201](http://html.mh-extra.com/html/waygate-technologies/krautkramer-usm-go/krautkramer-usm-go_operating-manual_878369201.webp)
Evaluation according to the DGS method
5 Operation
USM Go
Rev. 10 (06/2021)
5-125
– Select the function group
GATE A
.
– Select the function
TOF MODE
and press a function
key to set the measuring point in TOF mode to
PEAK
.
– Select the function group
AUTOCAL
and calibrate
the USM Go (see Chapter 5.8
Recording a reference echo and turning the
DGS curve on
To be able to display the required DGS curve, you have
to record the reference echo.
– Select the function group
DGS
.
– Use the navigation to move
upwards
in order to dis-
play the specific function groups and functions for the
DGS.
– Select the function group
GATE A
.
– Peak the echo of the reference reflector, in this case
the backwall echo from the test object.
– After this, position the gate A on the reference echo.
– Select the function group
REF ECHO
.
– Select the function
RECORD REF
and press a func-
tion key. The recording is confirmed by the indication
(recorded)
. The status icon
A DGS reference is
stored
is displayed in the top left corner, next to the
Содержание Krautkramer USM Go
Страница 1: ...Krautkrämer USM Go USM Go Operating Manual Id No 49 155 Rev 10 06 2021 ...
Страница 5: ...USM Go Rev 10 06 2021 0 5 Second operating level ...
Страница 6: ...0 6 Rev 10 06 2021 USM Go Second operating level continued ...
Страница 10: ...0 10 Rev 10 06 2021 USM Go ...
Страница 20: ...0 10 Rev 10 06 2021 USM Go ...
Страница 21: ...USM Go Rev 10 06 2021 1 1 Introduction 1 ...
Страница 35: ...USM Go Rev 10 06 2021 2 1 Standard package and accessories 2 ...
Страница 41: ...USM Go Rev 10 06 2021 3 1 Initial start up 3 ...
Страница 53: ...USM Go Rev 10 06 2021 4 1 Principles of operation 4 ...
Страница 76: ...4 Principles of operation Saving the settings 4 24 Rev 10 06 2021 USM Go ...
Страница 77: ...USM Go Rev 10 06 2021 5 1 Operation 5 ...
Страница 210: ...5 Operation Evaluation according to the DGS method 5 134 Rev 10 06 2021 USM Go ...
Страница 211: ...USM Go Rev 10 06 2021 6 1 Documentation 6 ...
Страница 240: ...6 30 Rev 10 06 2021 USM Go ...
Страница 241: ...USM Go Rev 10 06 2021 7 1 Maintenance and care 7 ...
Страница 247: ...USM Go Rev 10 06 2021 8 1 Interfaces and Peripherals 8 ...
Страница 251: ...USM Go Rev 10 06 2021 9 1 Appendix 9 ...
Страница 266: ...9 16 Rev 10 06 2021 USM Go ...
Страница 267: ...USM Go Rev 10 06 2021 10 1 Specifications 10 ...
Страница 278: ...10 12 Rev 10 06 2021 USM Go ...
Страница 279: ...USM Go Rev 10 06 2021 11 1 Index 11 ...