Wavelength References ClarityPlus Скачать руководство пользователя страница 1

WRD-19020-1 

 

 

Operator’s Manual 

 

ClarityPlus

TM

 

Full C Band Tunable Laser and 

Reference Standard 

Laser Locked to Molecular Absorption Lines 

 

 

Operating Manual WRD-19020-1 

 

Copyright 

 Wavelength References, 2019 

All Rights Reserved

 

 

 

Wavelength References 

1800 SW 3

rd

 Street, #140 

Corvallis, OR 97333 

Voice (541)738-0528 

Fax (541)-738-0998 

E-mail:  

[email protected]

 

 

 Wavelength References 

 

Содержание ClarityPlus

Страница 1: ...rd Laser Locked to Molecular Absorption Lines Operating Manual WRD 19020 1 Copyright Wavelength References 2019 All Rights Reserved Wavelength References 1800 SW 3rd Street 140 Corvallis OR 97333 Voic...

Страница 2: ...he product to Customer if the shipment is within the United States or Canada Customer shall be responsible for paying all shipping charges duties taxes and any other charges for products returned to o...

Страница 3: ...t Panel Operation and Displays 8 5 2 1 The Main Menu 8 5 2 2 The Mode Settings Display 8 5 2 3 The Mode Display 10 5 3 Front Panel Instrument Settings Display 11 5 3 1 Secure Laser Display 11 5 3 2 Ch...

Страница 4: ...e gas Occupational Safety and Health Administration OSHA lists a Permissable Exposure Limit PEL for HCN of 11mg m3 over an 8 hour period time weighted average The reference cell within the laser instr...

Страница 5: ...cial Publication 260 137 see section 7 1 and bibliography Furthermore the laser self calibrates to these reference lines allowing high accuracy wavelengths ANYWHERE in the C band The ClarityPlus ships...

Страница 6: ...Low noise area While the Clarity laser is designed for noise immunity the best performance will be achieved in a low EMF area Please use a noise line filter in an area where high noise is unavoidable...

Страница 7: ...e directional buttons to reach the desired new password press and hold for continuous scrolling Press ENTER then ENTER again to confirm or press MENU to cancel 5 1 2 Initial Warm Up When the user lock...

Страница 8: ...nu shown in Section 5 1 1 Button behavior is described below UP DOWN moves through the list of options ENTER goes to the appropriate sub menu specific mode settings or instrument settings MENU returns...

Страница 9: ...avelengths are available to the user but are not explicitly specified 5 2 2 3 Channel Frequency Lock Settings Display The Channel Frequency Lock Settings first highlights the Channel Spacing Channel R...

Страница 10: ...ettings display EXCEPT when laser is in Channel Frequency Lock Mode and locked In this case the user is able to update the offset without forcing a new calibration sequence See Section 5 2 3 4 for mor...

Страница 11: ...to enter the Channel Frequency Mode Settings display 5 3 Front Panel Instrument Settings Display The ClarityPlus includes access to additional instrument wide settings To access select Instrument Sett...

Страница 12: ...matically disable Power cycle instrument and try again waiting several seconds between power off and power on If problem persists consult factory 3 Drive Error Occurs when the drive current for the la...

Страница 13: ...annel Frequency Lock Optional enter target frequency in THz SOURce MODe FREQuency Queries lock mode Returns mode and target value SOURce MODe FREQuency RESolution 25 50 100 Sets Channel Frequency Lock...

Страница 14: ...ssword screen SYSTem SECure LASer Queries whether laser is in the password screen or not Returns ON or OFF SYSTem SERial BAUD 9600 19200 38400 5 7600 115200 Sets baud rate SYSTem SERial BAUD Queries b...

Страница 15: ...protected 220 Parameter error No entry in list to retrieve number list or channel list 221 Settings conflict Check rear interlock Settings Conflict Check rear interlock 222 Data out of range Data out...

Страница 16: ...that the ClarityPlus reports wavelength to the nearest 0 1pm Line Line ID Vacuum Wavelength nm Pressure Shift pm kPa Reference Cell Wavelength nm 0 R26 1527 633278 0 043 1527 63329 5 1 R25 1528 05458...

Страница 17: ...5555 4 36 P10 1549 730589 0 024 1549 73058 4 37 P11 1550 515625 0 050 1550 51561 4 38 P12 1551 310684 0 071 1551 31066 4 39 P13 1552 11576 0 091 1552 11573 5 40 P14 1552 93088 0 112 1552 93084 4 41 P1...

Страница 18: ...onal Information 8 1 Applications A typical application for the ClarityPlus laser is to provide an extremely accurate and stable wavelength standard in the DWDM band This can be used as a calibration...

Страница 19: ...extrapolated zero pressure line position Uncertainties are added in quadrature Pressure uncertainty in the gas fill also contributes to the line location and uncertainty though not by a measureable a...

Страница 20: ...ufactured with melted sealed glass to ensure long term stability of the gas cell environment The leak rate from this type of seal is extremely slow making the tubes effectively hermetically sealed goo...