CARDIO WAVE EXCITE Class & Trend: Service & maintenance manual- rev.2.0
Pagina 6.8
6.3.
THE DISPLAY FAILS TO ILLUMINATE
This error occurs when the power supply voltage is not reaching the upper assembly.
6.3.1.
500
AND
700
M
ODELS
Does the wall outlet supply the
correct voltage?
Is the mains lead OK?
Are the LEDs on the
Display board lit on?
Plug the machine into a
working electrical outlet
Replace the mains lead
Replace the Display
board
THE DISPLAY
DOES NOT
ILLUMINATE
NO
YES
NO
YES
NO
YES
A
1
Are the fuses OK?
Replace the fuses
YES
NO
Continued on the following page
Содержание CLASS & TREND
Страница 1: ...SERVICE MAINTENANCE MANUAL REV 2 0 ...
Страница 2: ......
Страница 4: ......
Страница 10: ...CARDIO WAVE EXCITE Class Trend Service maintenance manual rev 2 0 Pagina 1 6 Page intentionally left blank ...
Страница 32: ...CARDIO WAVE EXCITE Class Trend Service maintenance manual rev 2 0 Pagina 2 20 Page intentionally left blank ...
Страница 44: ...CARDIO WAVE EXCITE Class Trend Service maintenance manual rev 2 0 Pagina 3 12 Page intentionally left blank ...
Страница 50: ...CARDIO WAVE EXCITE Class Trend Service maintenance manual rev 2 0 Pagina 4 6 LIBYA DVB T ...
Страница 51: ...CARDIO WAVE EXCITE Class Trend Service maintenance manual rev 2 0 Pagina 4 7 Page intentionally left blank ...
Страница 52: ......
Страница 58: ...CARDIO WAVE EXCITE Class Trend Service maintenance manual rev 2 0 Pagina 5 6 Page intentionally left blank ...
Страница 100: ...CARDIO WAVE EXCITE Class Trend Service maintenance manual rev 2 0 Pagina 6 42 Page intentionally left blank ...
Страница 144: ...CARDIO WAVE EXCITE Class Trend Service maintenance manual rev 2 0 Pagina 7 44 Page intentionally left blank ...
Страница 202: ...CARDIO WAVE EXCITE Class Trend Service maintenance manual rev 2 0 Pagina 9 50 Page intentionally left blank ...
Страница 213: ......