10.4 HMI Cover
The HMI cover is illustrated by the picture below :
Follow the procedure below for the installation of the HMI cover
Procedure
1. Check the pump casing surrounding the HMI is clean and free of debris
2. Press the frame of the HMI cover onto the pump casing surrounding the HMI
3. Check the HMI cover flap lifts up and down freely without loosening the frame of the HMI cover
10.5 Other accessories
The procedure for the installation of other accessories such as the input/output control cables, hydraulic
connectors, are provided when relevant during further installation sections.
10 INSTALLATION: PART 1 (PHYSICAL)
54