Table 10 - Cyclic parameters
Index
Name
Access
Type
Description
1
SetFlowCal
Write
UInt32 Set the flow calibration value ( μL/rev)
2
SetSpeed
Write
UInt16
Speed is set in Deci RPM. Max speed
depends on head type. See Pump Head
enumeration table
3
SetSpeedLimit
Write
UInt16
Speed is set in Deci RPM. Max speed
depends on head type. See pump head
table below.
4
SetFailsafeSpeed
Write
UInt16
If the failsafe is enabled the pump will
run continuously at the selected speed in
the event of a communications loss.
5
SetFailsafeEnable
Write
Bool
Enabled the failsafe speed. If disabled
the pump will stop in the event of a
communications loss. If enabled the
pump will run at the speed set in the
“SetFailsafeSpeed” parameter
6
SetReverse
Write
Bool
If set the pump will run anti-clockwise.
Pump defaults to clockwise rotation
7
Run
Write
Bool
Set to 1 (true) to allow the pump to run. 0
will stop the pump. Note that pump
enable needs to be set
8
RunEnable
Write
Bool
Need to set to 1 to allow pump to run.
Setting to 0 will stop the pump and not
allow the pump to run.
9
ResetRunHours
Write
Bool
Resets the run hours accumulator
10
PauseFlowTotaliser
Write
Bool
Set to 1 to pause the internal
FlowTotaliser parameter. Setting to 0 will
un-pause the parameter
11
ResetFlowTotaliser
Write
Bool
Set to 1 to reset the flow totaliser. Set to
0 to allow the flow totaliser to
accumulate