14.2
First-time start-up defaults
The pump is preset with operational parameters as shown in table below.
Parameter
730 default
Language
Not set
Default mode
Manual
Default manual speed
360 rpm
Pump status
Stopped
Max speed
360 rpm
Direction
CW
Pumphead
720R
Tube size
25.4 mm
Tube material
Bioprene
Flow calibration
0.92 l/rev
Flow units
rpm
SG value
1
Keypad lock
Disabled
Beeper
ON
Security code
Not set
Remote start/stop input
High = stop
Leak detector input
High = leak
The pump is now ready to operate according to the defaults listed above.
Note: The display background colour changes according to running state as follows:
l
White background indicates pump stopped
l
Grey background indicates pump running
l
Red background indicates error or alarm
All operating parameters may be changed by means of key-presses (see section
).
m-730en-en-08.1
37