Installing and removing the corona pin
March 14, 2014, 715004387 Rev. B
93
Installing and removing the corona pin
For ESCi, APCI, and dual-mode APCI/APPI operation, you must fit a corona
pin.
Installing the corona pin in the source
Required material
Chemical-resistant, powder-free gloves
To install the corona pin in the source:
1. Prepare the instrument for working on the source (see
).
2. Pull the source enclosure release (located at the bottom, right-hand side)
outward, and swing open the enclosure.
Warning:
To avoid personal contamination with biologically
hazardous, or toxic materials, and to avoid spreading
contamination to uncontaminated surfaces, wear clean,
chemical-resistant, powder-free gloves while performing this
procedure. The LC system connections, ESI probe, and source can
be contaminated.
Warning:
To avoid electric shock, ensure that the instrument is
prepared for working on the source before commencing this
procedure.
Warning:
To avoid burn injuries, take great care while working
with the source enclosure open; the source can be hot.
Warning:
To avoid puncture wounds, take great care while
working with the source enclosure open if an ESI probe is fitted;
the ESI probe tip is sharp.
Содержание Xevo TQD
Страница 2: ...ii March 14 2014 715004387 Rev B...
Страница 22: ...xxii March 14 2014 715004387 Rev B...
Страница 56: ...2 Preparing for Operation 56 March 14 2014 715004387 Rev B...
Страница 200: ...4 Maintenance Procedures 200 March 14 2014 715004387 Rev B...
Страница 216: ...A Safety Advisories 216 March 14 2014 715004387 Rev B...
Страница 242: ...B External Connections 242 March 14 2014 715004387 Rev B...
Страница 246: ...C Materials of Construction and Compliant Solvents 246 March 14 2014 715004387 Rev B...