Spare parts
March 14, 2014, 715004387 Rev. B
85
Spare parts
Replace only spare parts, which are the parts mentioned in this document. For
details about spare parts, use the Waters Quality Parts
®
Locator on the
Waters Web site’s Services & Support page.
Replace the roughing pump’s
demister element.
Annually.
Tip:
Applications that
contaminate the
roughing pump oil
reduce this period,
which must be
determined from
experience.
.
Clean the APPI/APCI source
UV lamp window.
When the window
becomes visibly dirty
or when the sensitivity
decreases to
unacceptable levels.
.
Change the APPI/APCI source
UV lamp bulb.
When the bulb fails.
.
Replace the APPI lamp drive
assembly O-rings.
Annually.
See
.
Maintenance schedule:
Procedure
Frequency
For information...