NanoFlow source
March 14, 2014, 715004387 Rev. B
77
Fitting a borosilicate glass capillary (nanovial)
Required materials
•
Chemical-resistant, powder-free gloves
•
Needle-nose pliers
•
Borosilicate glass capillary
•
Fused silica syringe needle or GELoader
®
tip
•
Fused silica cutter
To fit a borosilicate glass capillary (nanovial):
1. Loosen the stage retaining screw.
2. Pull the stop screw to release the stage.
3. Slide the stage out of the NanoFlow source enclosure and remove the
magnetic cover.
4. Unscrew the retaining screw, and lift the sprayer from the stage.
Warning:
To avoid lacerations, puncture injuries, and
possible contamination with biohazardous and toxic samples,
do not touch the sharp end of the capillary.
Notice:
To avoid damaging capillaries, take great care when
handling them. Capillaries are extremely fragile; always
hold their blunt end, never the sharp end, which can easily
be damaged.
Warning:
To avoid electric shock, ensure that the NanoFlow
stage is fully retracted from the source before beginning this
procedure.
Содержание Xevo TQD
Страница 2: ...ii March 14 2014 715004387 Rev B...
Страница 22: ...xxii March 14 2014 715004387 Rev B...
Страница 56: ...2 Preparing for Operation 56 March 14 2014 715004387 Rev B...
Страница 200: ...4 Maintenance Procedures 200 March 14 2014 715004387 Rev B...
Страница 216: ...A Safety Advisories 216 March 14 2014 715004387 Rev B...
Страница 242: ...B External Connections 242 March 14 2014 715004387 Rev B...
Страница 246: ...C Materials of Construction and Compliant Solvents 246 March 14 2014 715004387 Rev B...