B External Connections
232
March 14, 2014, 715004387 Rev. B
6. Use snoop (or equivalent) leak detector liquid to ensure that there are no
leaks at the instrument exhaust and laboratory exhaust system line
connections.
Exhaust trap bottle:
Warning:
To prevent leaks in the source exhaust system,
confirm the integrity of the system by performing the
following leak test. Leaks can result in the release of
biohazardous or toxic materials.
Notice:
To avoid damage to the instrument, use snoop (or its
equivalent) leak detector liquid only for the purpose
described in the following step. Do not use it on any other
part of the instrument.
From instrument
waste (12-mm)
To laboratory
exhaust (12-mm)
Valve control
cable (from
instrument)