4.6.2 Fitting the source enclosure to the instrument
Required materials
Chemical-resistant, powder-free gloves
To fit the source enclosure to the instrument:
Warning:
To avoid personal contamination with biohazards, toxic materials, and corrosive
materials, wear chemical-resistant gloves when performing this procedure.
Warning:
To avoid puncture injuries, handle sample needles, syringes, fused silica
lines, and borosilicate tips with extreme care.
1.
Using both hands, fit the source enclosure to the two supporting studs on the source
adaptor housing.
Notice:
To avoid damaging the sample inlet, when removing a NanoLockSpray
source enclosure, you must slide the sprayer platform out of the source enclosure
before you open the enclosure.
2.
Close the source enclosure.
3.
Connect the probe adjuster and options cables to the instrument’s connectors.
Tip:
The cables and connectors are color coded; the blue-sleeved cable connects to the
blue connector and the yellow-sleeved cable to the yellow connector.
4.
Slide closed the instrument’s source interface door.
4.7 Operating the source isolation valve
You must close the source isolation valve to isolate the source from the instrument vacuum
system for certain maintenance procedures.
Required materials
Chemical-resistant, powder-free gloves
June 9, 2016, 715004990 Rev. A
Page 98