Cleaning the sampling cone assembly
4-33
2. Carefully fit the sample cone into the cone gas nozzle.
3. Fit the O-ring into the groove created between the sample cone and cone
gas nozzle. (Fit a new O-ring if the old one has been disposed of.)
Fitting the sampling cone assembly to the source
Required material
Chemical-resistant, powder-free gloves
To fit the sampling cone assembly to the source
1. Ensure that the source isolation valve is in the closed position (see
close the source isolation valve before starting a maintenance procedure”
).
Warning:
The source components can be contaminated with
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
Warning:
To avoid puncture wounds, take great care while working
with the source enclosure open, because the ESI probe tip is sharp.
Caution:
To avoid damage, do not open the source isolation valve
before fitting the sampling cone assembly to the ion block assembly.