4-28
Maintenance Procedures
Disassembling the sampling cone assembly
Required material
•
Chemical-resistant, powder-free gloves
•
Combined 2.5-mm Allen wrench and cone extraction tool
To disassemble the sampling cone assembly
1. On the combined 2.5-mm Allen wrench and cone extraction tool, slide
the collar to the end of the tool.
2. Insert the collar in the sample cone.
Warning:
The sampling cone assembly can be contaminated with
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
Collar