4-14
Maintenance Procedures
Removing O-rings and seals
When performing certain maintenance procedures, you must remove O-rings
or seals from instrument components. An O-ring removal kit is provided with
the instrument.
O-ring removal kit
To remove an O-ring
1. Use the forked end of tool 1 to impale the O-ring or seal.
2. Pull the O-ring or seal from its groove; if necessary, use tool 2 as an aid.
3. Dispose of the O-ring or seal in accordance with local environmental
regulations.
Caution:
When removing an O-ring or seal from a component, be
careful not to scratch the component with the removal tool.
Warning:
The O-ring or seal can be contaminated with
biohazardous and/or toxic materials. Dispose of it according
to local environmental regulations.