Preparing the IntelliStart Fluidics system
2-5
Running the mass spectrometer at high flow rates
The ACQUITY UPLC
®
system runs at high flow rates. To optimize
desolvation, and thus sensitivity, run the ACQUITY
®
Xevo TQ MS system at
appropriate gas flows and desolvation temperatures. IntelliStart software
automatically sets these parameters when you enter a flow rate, according to
the following table.
Preparing the IntelliStart Fluidics system
For additional information, see
“Connecting the liquid waste line” on
Installing the reservoir bottles
Use standard reservoir bottles (30 mL) for instrument setup and calibration.
Use the Low-volume Adaptor Kit (included) to infuse smaller volumes. The
low-volume vials have a volume of 1.5 mL.
Required materials
Chemical-resistant, powder-free gloves
Flow rate versus temperature and gas flow
Flow rate
(mL/min)
Source temp (°C)
Desolvation
temp (°C)
Desolvation gas
flow (L/h)
0.000 to 0.020
150
200
800
0.020 to 0.100
150
300
800
0.101 to 0.300
150
400
800
0.301 to 0.500
150
500
1000
> 0.500
150
0
1000
Caution:
To avoid accidental spillage damaging the instrument, do not
store large volume solvent reservoirs on top of the instrument.