Replacing the source assembly seals
4-75
Replacing the source assembly seals
To avoid excessive leakage of solvent vapor into the laboratory atmosphere,
the following seals must be renewed at intervals of no greater than 1 year:
•
Probe adjuster assembly probe seal
•
Probe adjuster assembly nebulization gas seal
•
Source enclosure seal
•
Source enclosure nebulizer gas seal
•
Source enclosure desolvation gas seal
•
Source enclosure pressure test gas seal
Removing the probe adjuster assembly probe and source
enclosure seals
Required materials
•
Chemical-resistant, powder-free gloves
•
O-ring removal kit
Warning:
To avoid excessive leakage of solvent vapor into the
laboratory atmosphere, the seals listed below must be renewed,
at intervals of no greater than 1 year, exactly as described in this
section.