Replacing the ESI probe tip and gasket
4-59
Replacing the ESI probe tip and gasket
Replace the ESI probe tip if a blockage occurs in the internal metal sheathing
through which the stainless steel capillary passes or if the probe tip is
damaged.
Removing the ESI probe tip and gasket
Required materials
•
Chemical-resistant, powder-free gloves
•
7-mm wrench
•
10-mm wrench
To remove the ESI probe tip and gasket
1. Remove the ESI probe from the source (see
).
Warning:
The probe and source components can be contaminated
with biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
Warning:
The probe and source can be hot. To avoid burn injuries, take
great care while performing this procedure.
Warning:
The ESI probe tip is sharp. To avoid puncture wounds,
handle the probe with care.