Configuring for APCI mode
3-13
24. Use the combined 2.5-mm Allen wrench to fit and tighten the 3 probe
end-cover retaining screws.
25. Fit the ESI probe to the source (see
).
Configuring for APCI mode
To operate in APCI mode, you must fit the APCI probe and corona pin to the
LockSpray source enclosure.
For more information on using APCI mode, see the Xevo G2 QTof system
online Help.
Installing the APCI probe
Required materials
•
Chemical-resistant, powder-free gloves
•
Sharp knife or PEEK tubing cutter
End-cover retaining screws