Installing the small-bore capillary option
3-7
Installing the small-bore capillary option
Use the small-bore capillary option with 300-µm UPLC columns in systems
running at typical flow rates of 10 µL/min. The materials necessary to install
the option are supplied in kit form.
Required materials
•
Chemical-resistant, powder-free gloves
•
Combined 2.5-mm Allen wrench and cone extraction tool
•
10-mm wrench
•
8-mm wrench
•
7-mm wrenches (2)
•
LC pump
•
HPLC-grade (or better) 1:1 acetonitrile/water
•
Sharp knife or PEEK tubing cutter
•
From the small-bore capillary kit
–
capillary
–
small-bore, UNF coupler (slide port)
–
collar nut (thumb nut)
–
PTFE liner tubing
–
conductive sleeve
–
two ferrules (1/16-inch)
•
Metal gasket for the probe tip
•
Red PEEK tubing
Caution:
To avoid damage from excessive pressure, do not exceed flow
rates of 50 µL/min through the ESI probe when the small-bore
capillary is fitted.