Starting the mass spectrometer
2-3
6. Allow 4 minutes for the embedded PC to initialize.
Tip:
The power and status LEDs change as follows:
•
During initialization, the binary solvent manager’s and sample
manager’s status LED flashes green.
•
After the instruments are successfully powered-on, all power LEDs
show steady green. The binary solvent manager’s flow LED, the
sample manager’s run LED, and the mass spectrometer’s status
LED remain unlit.
7. Start the software and monitor the instrument console for messages and
LED indications.
8. Click Operate.
Result:
When the mass spectrometer is in good operating condition, the
software indicates “Ready” in the instrument console.
Verifying the instrument’s state of readiness
When the mass spectrometer is in good operating condition, the power and
status LEDs show constant green. You can view any error messages in console
software.
Monitoring the mass spectrometer LEDs
Light-emitting diodes on the mass spectrometer indicate its operational
status.
Power LED – The power LED, below the mass spectrometer’s source, indicates
when the mass spectrometer is powered-on or powered-off.
Status LED – The status LED, on the right-hand side of the power LED,
indicates the operating condition. See the mass spectrometer’s online Help for
details on the status LED indications.
Calibration
Calibrate the mass spectrometer prior to use, see the mass spectrometer’s
online Help.
Содержание Xevo G2 QTof
Страница 10: ...x...
Страница 18: ...xviii Table of Contents...
Страница 34: ...1 16 Waters Xevo G2 QTof Overview...
Страница 44: ...2 10 Preparing the Mass Spectrometer for Operation...
Страница 108: ...5 32 Maintenance Procedures 2 Slide the collar to the end of the tool 3 Insert the collar in the sample cone Collar...
Страница 227: ...Connecting the liquid waste line B 17 Positioning of drain tube TP01807 Correct Incorrect...
Страница 238: ...B 28 External Connections...
Страница 242: ...C 4 Materials of Construction and Compatible Solvents...
Страница 248: ...Index 6...