Connecting the Varian oil-filled roughing pump
B-3
Connecting the Varian oil-filled roughing pump
Note:
To connect the alternative oil-free roughing pump, see
.
Required materials
•
Chemical-resistant, powder-free gloves
•
7-mm nut driver
•
Sharp knife
•
NW25 center rings (included in the Xevo™ G2 QTof installation kit)
•
NW25 clamps (included in the Xevo G2 QTof installation kit)
•
12.7-mm clear PVC exhaust tubing (included in the Xevo G2 QTof
installation kit)
•
PVC hose clamps (included in the Xevo G2 QTof installation kit)
•
Vacuum hose (included in the Xevo G2 QTof installation kit)
•
PTFE drip tray
TP03159
Inlet port flange
Oil exhaust filter
Electrical connection
panel