5-34
Maintenance Procedures
Cleaning the sample cone and cone gas nozzle
Required materials
•
Chemical-resistant, powder-free gloves.
•
Appropriately sized glass vessels in which to completely immerse
components when cleaning. Use only glassware not previously cleaned
with surfactants.
•
HPLC-grade (or better) methanol.
•
HPLC-grade (or better) water.
•
Formic acid.
•
Ultrasonic bath.
•
Source of oil-free, inert gas (nitrogen or argon) for drying (air-drying
optional).
•
Wash bottle containing HPLC-grade (or better) 1:1 methanol/water.
•
Large beaker.
To clean the sample cone and cone gas nozzle:
1. If the sample cone contains debris, place a drop of formic acid on its
orifice.
2. Immerse the sample cone, cone gas nozzle, and cone gas nozzle handle in
separate glass vessels containing 1:1 methanol/water.
Tip:
If the components are obviously contaminated, use 45:45:10
methanol/water/formic acid.
3. Place the vessels in the ultrasonic bath for 30 minutes.
Warning:
The sample cone and cone gas nozzle can be
contaminated with biohazardous and/or toxic materials. Always
wear chemical-resistant, powder-free gloves while performing this
procedure.
Warning:
Formic acid is extremely corrosive and toxic. Work with
extreme care. Use a fume hood and suitable protective equipment.
Caution:
The sample cone is fragile. Never place it on its tip; always
place it on its flanged base.
Содержание Xevo G2 QTof
Страница 10: ...x...
Страница 18: ...xviii Table of Contents...
Страница 34: ...1 16 Waters Xevo G2 QTof Overview...
Страница 44: ...2 10 Preparing the Mass Spectrometer for Operation...
Страница 108: ...5 32 Maintenance Procedures 2 Slide the collar to the end of the tool 3 Insert the collar in the sample cone Collar...
Страница 227: ...Connecting the liquid waste line B 17 Positioning of drain tube TP01807 Correct Incorrect...
Страница 238: ...B 28 External Connections...
Страница 242: ...C 4 Materials of Construction and Compatible Solvents...
Страница 248: ...Index 6...