Cleaning the sampling cone assembly
5-31
Disassembling the sampling cone assembly
Required materials
•
Chemical-resistant, powder-free gloves
•
Combined 2.5-mm Allen wrench and cone extraction tool
To disassemble the sampling cone assembly:
1. Retrieve the combined 2.5-mm Allen wrench and cone extraction tool
from its storage location behind the source enclosure.
Warning:
The sampling cone assembly can be contaminated with
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
TP03130
Combined 2.5-mm
Allen wrench and cone
extraction tool