Shutting down the TQ Detector
2-15
Shutting down the TQ Detector
Recommendation:
You should always leave the TQ Detector in Operate mode
except when performing routine maintenance. It is not necessary to switch to
Standby mode. However, shutting down the TQ Detector is acceptable
provided that you consider warm-up time on restarting. If you do shut down
the TQ Detector, refer to the instructions in this section.
Tip:
Set system shutdown parameters in the Shutdown Editor. Consult the
MassLynx online Help for more information.
Putting the TQ Detector in Standby mode for overnight shutdown
You might want to shut down the instrument for a relatively brief period, like
overnight, while maintaining the LC flow.
To put the TQ Detector in Standby mode:
1.
In the ACQUITY UPLC Console, click
to stop the LC flow or, if
column flow is required, divert the LC flow to waste as follows:
a.
In the ACQUITY UPLC Console system tree, expand TQ Detector.
b.
Click Interactive Display.
c.
Click the current diverter valve position label.
d.
In the Select a Flow State dialog box, select Waste.
e.
Click OK.
2.
Click Standby
to shut off heaters and voltages.
Tip:
You can create a method to stop the gas flow or lower the
temperature. See the MassLynx online Help for more information on
creating methods.
Caution:
Buffers left in the system can precipitate and damage
instrument components.