3-18
Configuring the LockSpray Source
Installing the corona pin in the source
Install the corona pin according to the procedure on
Removing the corona pin from the source
Remove the corona pin according to the procedure on
Removing the APCI probe
Required materials
Chemical-resistant, powder-free gloves
To remove the APCI probe
1. Prepare the instrument for work performed on its source (see
).
2. Disconnect the diverter valve tubing from the APCI probe.
3. Unscrew the probe locking ring.
4. Carefully remove the probe from the probe adjuster assembly.
Warning:
The LC system connections, APCI probe, and source can
be contaminated with biohazardous and/or toxic materials. Always
wear chemical-resistant, powder-free gloves while performing this
procedure.
Warning:
To avoid electric shock, prepare the instrument for work
performed on its source before commencing this procedure.