C-2
Materials of construction and compliant solvents
Preventing contamination
For information on preventing contamination, see
Controlling Contamination
in LC/MS Systems
(part number 715001307). You can find this document on
http://www.waters.com; click Services and Support and then Support Center.
Items exposed to solvent
The items that appear in the following table can be exposed to solvent. You
must evaluate the safety issues if the solvents used in your application differ
from the solvents normally used with these items. See
for details
about the most common ingredients used to prepare mobile phases.]
Items exposed to solvent
Item
Material
Corona discharge pin mounting
contact
PEEK™
a
Gas exhaust port
Aluminium
Gas tubes
FEP
b
Ion block
Stainless steel
Ion block support
PEEK
a
Isolation valve
Gold-plated aluminium/bronze
O-rings
Viton
®
or PTFE
c
-encapsulated Viton
Probe adjuster bellows
PTFE
c
/Viton
Probe adjuster assembly
Anodized aluminium, glass filled
acetal, and stainless steel
Probe shaft
PEEK
a
Push-in gas fittings
Nickel/brass
Solvent waste/leak management
Tygon tubing
Source enclosure
Alochromed aluminium
Source enclosure view port
Toughened plate glass
Waste bottle
Polypropylene
Waste bottle push-in fittings
NBR
d
, SST
e
, PBT
f
, and POM
g
Содержание SYNAPT G2
Страница 18: ...xviii Table of Contents...
Страница 46: ...2 8 Starting Up and Shutting Down the Mass Spectrometer...
Страница 66: ...3 20 Configuring the LockSpray Source...
Страница 228: ...B 24 External Connections 7 Click Next Result The software installs 8 Click Finish...
Страница 232: ...C 4 Materials of construction and compliant solvents...