Replacing the LockSpray source assembly seals
5-93
4. Dispose of all the seals in accordance with local environmental
regulations.
Fitting the new source enclosure seals
Required materials
•
Chemical-resistant, powder-free gloves
•
Wash bottle containing HPLC-grade (or better) 1:1 methanol/water
•
New seals
Warning:
The seals can be contaminated with biohazardous
and/or toxic materials. Dispose of them according to local
environmental regulations.
Source enclosure seal
Nebulizer gas seal
Desolvation gas
seal