Warning:
To avoid personal contamination with biohazardous or toxic compounds, wear clean,
chemical-resistant, powder-free gloves when performing this procedure.
Warning:
To prevent personal injury, always observe Good Laboratory Practice when
handling solvents, changing tubing, or operating the detector. Know the physical and
chemical properties of the solvents used (see the Material Safety Data Sheets for the
solvents in use).
Warning:
To avoid electric shock, observe these precautions:
• Do not remove the mass spectrometer’s protective panels. The components they
cover are not user-serviceable.
• When the instrument is in Operate mode, avoid touching the areas marked with the
high voltage warning symbol. To touch external areas marked with the symbol, first
put the instrument in Standby mode.
Warning:
To avoid burn injuries, take great care while working with the probe and
source; these components can be hot.
Warning:
To avoid injury, ensure that these criteria are met when performing
maintenance operations inside the source enclosure:
• The instrument is in Standby mode.
• LC flow is diverted to waste or set to Off.
• Desolvation gas flow is stopped.
See
for safety advisory information.
3.7 Removing and refitting the source enclosure
Remove the source enclosure to gain access to the source components that need routine
cleaning and replacement.
3.7.1 Removing the source enclosure from the instrument
Warning:
To avoid personal contamination with biohazardous or toxic compounds, wear clean,
chemical-resistant, powder-free gloves when performing this procedure.
Notice:
To avoid damaging the fragile probe, use care when removing it from the
source enclosure.
June 15, 2017, 715003956 Rev. F
Page 28