Figure 3–90: Support-plate bolts on accumulator pump head
1
Support-plate bolts
Figure 3–91: Removing accumulator pump head and support plate from actuator
housing
Support plate
Pump head
Plunger
8.
Stand the pump head upright on a clean surface.
3.19.3 Removing the pump head plunger
Warning:
Observe Good Laboratory Practice (GLP) at all times, particularly when
working with hazardous materials. Consult the Material Safety Data Sheets regarding
the solvents you use. Additionally, consult the safety representative for your
organization regarding its protocols for handling such materials.
Notice:
To avoid contaminating system components, wear clean, chemical-resistant,
powder-free gloves when performing this procedure.
August 8, 2016, 715005050 Rev. C