Column heater set point
– Displays the current column heater set point, to 0.1 °C
resolution. When active temperature control is disabled, this field displays “Off”.
Sample compartment set point
– Displays the current sample compartment set point, to
0.1 °C resolution. When active temperature control is disabled, this field displays “Off”.
Current sample compartment temperature
– Displays the current sample compartment
temperature, to 0.1 °C resolution, even when active temperature control is disabled.
Status
– Displays the status of the current operation.
You can access the following additional functions by right-clicking anywhere in the sample
manager control panel:
Table 3–5: Additional functions in the sample manager control panel
Control panel function
Description
Prime syringes
Displays the Prime dialog box.
Wash needle
Displays the Wash Needle dialog box.
Turn lights on
Illuminates the ACQUITY UPLC Console.
Launch Console
Launches the ACQUITY UPLC Console.
Reset SM
Resets the sample manager following an error condition.
Help
Displays the ACQUITY UPLC Console online Help.
3.5.3 TUV control panel
The tunable ultraviolet (TUV) detector's control panel displays absorbance units and wavelength
values. When the detector is running in dual mode, the values for both wavelength A and B
appear.
You can edit detector parameters when the system is idle by clicking the underlined value. You
cannot edit these values while the system is running samples.
Figure 3–3: TUV detector control panel
The following table describes the items in the TUV’s control panel:
April 4, 2018, 715005704 Rev. A