34
Shutting down the detector
To shut down the detector:
1. Power-off the detector.
2. Stop the solvent flow to vent the system:
a. From the console, select Binary Solvent Manager in the system tree.
b. Enter zero flow in the composition field or press the Stop Flow icon
and enter a flow rate of zero. The system will vent to atmospheric
pressure through the vent valve in the ACQUITY UPC
2
Convergence Manager.
Maintaining the detector
Contacting Waters technical service
If you are located in the USA or Canada, report malfunctions or other
problems to Waters Technical Service (800 252-4752). If you are located
elsewhere, phone the Waters corporate headquarters in Milford,
Massachusetts (USA), or contact your local Waters subsidiary. Our Web site
includes phone numbers and e-mail addresses for Waters locations worldwide.
Go to www.waters.com, and click Waters Division.
When you contact Waters, be prepared to provide this information:
•
A record of error messages, if any
•
Nature of the symptom
•
Instrument serial numbers
•
Flow rate
•
Operating pressure
Warning:
To avoid burn injuries resulting from skin contact with
pressurized, liquid CO
2
and breathing difficulties caused by the
displacement of oxygen within the confines of the laboratory space,
stop the solvent flow, and vent the system pressure before you
attempt to loosen any fittings or disconnect the flow cell.