Starting the detector
29
8. Press the power switch on the front panel to power-on the detector.
Result:
The detector runs a series of startup diagnostic tests while the
lamp LED blinks green. The lamp LED shows steady green when the
lamp is ignited.
9. When the lamp LED is steady green, download an instrument or inlet
method in the Empower or MassLynx software.
Result:
The console displays messages and visual signals.
10. For best results, wait one hour for the detector to stabilize before
acquiring data.
Monitoring detector LEDs
Light emitting diodes on the detector indicate its state of functioning.
Power LED
The power LED, to the left-hand side of the detector’s front panel, indicates
when the detector is powered-on or powered-off.
Lamp LED
The lamp LED, to the right of the power LED, indicates the lamp status.
Lamp LED indications:
LED mode and color
Description
Unlit
Indicates the detector lamp is extinguished.
Constant green
Indicates the detector lamp is ignited.
Flashing Green
Indicates the detector is initializing or
calibrating.
Flashing red
Indicates an error stopped the detector.
Information regarding the error that caused the
failure can be found in the console.
Constant red
Indicates a detector failure that prevents further
operation. Power-off the detector, and then
power-on. If the LED is still steady red, contact
your Waters service representative.