June 28, 2016, 715003956 Rev. C
Page 35
5.
Remove the ion block from the source housing.
3.7.4
Cleaning the source components
Required materials:
•
Chemical-resistant, powder-free gloves.
•
Appropriately sized glass vessels in which to completely immerse components when
cleaning. Use only glassware not previously cleaned with surfactants.
•
HPLC-grade (or better) methanol.
•
HPLC-grade (or better) water.
•
Formic acid.
•
Ultrasonic bath.
•
Source of oil-free, inert gas (nitrogen or argon) for drying (air-drying optional).
•
Wash bottle containing HPLC-grade (or better) 1:1 methanol/water.
•
Large beaker.
To maintain the source components:
1.
Remove any O-rings from the source components, and replace the PEEK calibration pin at
the top of the ion block.
Important:
If the O-ring is visibly damaged, replace it.
Warning:
To avoid personal contamination with biohazards or toxic materials, always wear
chemical-resistant, powder-free gloves while performing this procedure.
Warning:
To avoid contamination with corrosive and toxic formic acid, use a fume hood and
suitable protective equipment during this procedure.