Audience and purpose
This guide is intended for use by individuals who install, operate, and
maintain the ACQUITY
®
30-cm Single-Zone Column Manager (CM-30S). It
gives an overview of the technology and operation of the CM-30S.
Intended use of the 30-cm single-zone column manager
Waters designed the 30-cm, single zone column manager for use in Advanced
Polymer Chromatography
™
(APC
™
) and Ultraperformance Convergence
Chromatography
™
(UPC
2
®) applications. The CM-30S is not intended for use
in diagnostic applications.
Calibrating
To calibrate LC systems, follow acceptable calibration methods using at least
five standards to generate a standard curve. The concentration range for
standards must include the entire range of QC samples, typical specimens,
and atypical specimens.
Quality control
Routinely run three QC samples that represent subnormal, normal, and
above-normal levels of a compound. If sample trays are the same or very
similar, vary the location of the QC samples in the trays. Ensure that QC
sample results fall within an acceptable range, and evaluate precision from
day to day and run to run. Data collected when QC samples are out of range
might not be valid. Do not report these data until you are certain that the
instrument performs satisfactorily.
Part number catalog number
Symbol
Definition
5()
Содержание ACQUITY CM-30S
Страница 2: ...ii October 7 2014 715004178 Rev B...
Страница 10: ...x October 7 2014 715004178 Rev B...
Страница 14: ...xiv October 7 2014 715004178 Rev B...
Страница 126: ...3 Maintaining the 30 cm single zone column manager 126 October 7 2014 715004178 Rev B...
Страница 138: ...A Safety Advisories 138 October 7 2014 715004178 Rev B...
Страница 146: ...B Specifications 146 October 7 2014 715004178 Rev B...