Installation recommendations for fittings
October 7, 2014, 715004178 Rev. B
41
You can access additional functions by right-clicking anywhere in the CM-30S
control panel.
Installation recommendations for fittings
The system uses gold-plated compression screws and two-piece ferrules. See
the diagram below for assembly orientation.
Recommendations:
•
To prevent bandspreading, ensure the tubing bottoms in its fitting hole
before tightening the compression screw.
•
For easier accessibility, use long compression screws to attach tubes to
the injector and vent valve.
•
Whenever you loosen fittings during maintenance, examine them for
cracks, stripped threads, and deformations.
•
Do not reuse stainless steel fittings more than six times.
Additional functions in the CM-30S control panel:
Control panel function
Description
Launch instrument console
Launches the instrument console.
Reset CM
Resets the CM-30S after an error
condition.
Help
Displays the online Help for the
instrument console software.
Warning:
To avoid the harmful effects of personal contact with
fittings contaminated with biologically hazardous or toxic
materials, always wear clean, chemical-resistant, powder-free
gloves when reinstalling the fittings.
Tubing
Compression screw
Ferrule with
locking ring
Содержание ACQUITY CM-30S
Страница 2: ...ii October 7 2014 715004178 Rev B...
Страница 10: ...x October 7 2014 715004178 Rev B...
Страница 14: ...xiv October 7 2014 715004178 Rev B...
Страница 126: ...3 Maintaining the 30 cm single zone column manager 126 October 7 2014 715004178 Rev B...
Страница 138: ...A Safety Advisories 138 October 7 2014 715004178 Rev B...
Страница 146: ...B Specifications 146 October 7 2014 715004178 Rev B...